Invacare Oxygen Equipment IRC5PO2V User Manual

User Manual  
Perfecto2Series  
Oxygen Concentrators without SensO2®,  
Oxygen Concentrators with SensO2  
HomeFill®Compatible  
Perfecto2 with SensO2  
Perfecto2  
Model IRC5PO2  
Model IRC5PO2W  
Model IRC5PO2V  
Model IRC5P  
Model IRC5PW  
Dealer: This manual MUST be given to the end  
user of this product.  
User: BEFORE using this product, read this  
manual and save for future reference.  
For more information regarding  
Invacare products, parts, and services,  
 
 
Table of Contents  
TABLE OF CONTENTS  
To Reduce the Risk of Burns, Electrocution, Fire or Injury to  
PARAMETERS ............................................... 12  
3
Perfecto2Series  
 
SPECIAL NOTES  
SPECIAL NOTES  
Signal words are used in this manual and apply to hazards or  
unsafe practices which could result in personal injury or  
property damage. Refer to the following table for definitions  
of the signal words.  
SIGNAL WORD  
MEANING  
Danger indicates an imminently hazardous situation  
which, if not avoided, will result in death or serious  
injury.  
DANGER  
Warning indicates a potentially hazardous situation  
which, if not avoided, could result in death or  
serious injury.  
WARNING  
Caution indicates a potentially hazardous situation  
which, if not avoided, may result in property  
damage or minor injury or both.  
CAUTION  
NOTICE  
The information contained in this document is subject to  
change without notice.  
DANGER  
DO NOT SMOKE while using this device. Keep all  
matches, lighted cigarettes or other sources of ignition  
out of the room in which this product is located and away  
from where oxygen is being delivered.  
NO SMOKING signs should be prominently displayed.  
Textiles and other materials that normally would not  
burn are easily ignited and burn with great intensity in  
oxygen enriched air. Failure to observe this warning can  
result in severe fire, property damage and cause physical  
injury or death.  
Perfecto2Series  
4
 
 
SPECIAL NOTES  
CAUTION  
“Caution: Federal law restricts this device to sale or rental by  
or on order of a physician, or any other practitioner licensed  
by the law of the State in which he/she practices to use or  
order the use of this device.”  
Invacare recommends an alternate source of supplemental  
oxygen in the event of a power outage, alarm condition or  
mechanical failure. Consult your physician or equipment  
provider for the type of reserve system required.  
This equipment is to be used as an oxygen supplement and is  
not considered life supporting or life sustaining.  
5
Perfecto2Series  
 
LABEL LOCATION  
LABEL LOCATION  
Perfecto2™  
FOR OPERATING INSTRUCTIONS or ALARMS (See Operator’s Manual)  
DANGER - RISK OF FIRE  
NO SMOKING - Keep ALL sources of ignition out of the room in which this product is located and  
away from areas where oxygen is being delivered. Textiles, oil and other combustibles are easily  
ignited and burn with great intensity in oxygen enriched air.  
DANGER - ELECTRIC SHOCK  
DO NOT remove cover. Refer servicing to qualified service personnel.  
Flowrates below 0.5 L/min. will trigger Low Flow Alarm (rapid audible alarm).  
Set flowmeter at 3.0 L/min. or less when filling HomeFill® cylinders.  
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician.  
HomeFill®  
Compatible  
Serial Number Label is  
located on the resonator  
intake assembly  
Specification Label is  
located on the back  
near the bottom  
Perfecto2Series  
6
 
 
SECTION 1—GENERAL GUIDELINES  
SECTION 1—GENERAL GUIDELINES  
In order to ensure the safe installation, assembly and  
operation of the Perfecto2 concentrator these instructions  
MUST be followed.  
WARNING  
This section contains important information for the safe  
operation and use of this product.  
To Reduce the Risk of Burns,  
Electrocution, Fire or Injury to Persons  
DANGER  
Risk of electric shock. DO NOT disassemble. Refer servicing  
to qualified service personnel. No user serviceable parts.  
Avoid using while bathing. If continuous usage is required  
by the physician’s prescription, the concentrator MUST be  
located in another room at least 7 ft from the bath.  
DO NOT come in contact with the concentrator while wet.  
DO NOT place or store product where it can drop into  
water or other liquid.  
DO NOT reach for product that has fallen into water.  
Unplug IMMEDIATELY.  
If the concentrator has a damaged cord or plug, if it is not  
working properly, if it has been dropped or damaged, or  
dropped into water, call qualified technician for  
examination and repair.  
A spontaneous and violent ignition may occur if oil, grease  
or greasy substances come in contact with oxygen under  
pressure. These substances MUST be kept away from the  
oxygen concentrator, tubing and connections, and all  
other oxygen equipment. DO NOT use any lubricants  
unless recommended by Invacare.  
Avoid creation of any spark near medical oxygen  
equipment. This includes sparks from static electricity  
created by any type of friction.  
7
Perfecto2Series  
 
   
SECTION 1—GENERAL GUIDELINES  
Operating Information  
For optimum performance, Invacare recommends that each  
concentrator be on and running for a minimum of 30 minutes  
at a time. Shorter periods of operation may reduce maximum  
product life.  
Keep the oxygen tubing, cord, and unit out from under such  
items as blankets, bed coverings, chair cushions, clothing and  
away from heated or hot surfaces, including space heaters,  
stoves and similar electrical appliances.  
DO NOT move or relocate concentrator by pulling on the  
power cord.  
NEVER drop or insert any object or liquid into any opening.  
Invacare recommends that CrushProof oxygen tubing be  
used with this product and not exceed 50 ft in length.  
There are no user serviceable parts. This does not include  
normal maintenance items. See maintenance section for user  
maintenance items.  
A product should NEVER be left unattended when plugged  
in. Make sure the Perfecto2 is Off when not in use.  
Close supervision is necessary when this product is used near  
children or physicallychallenged individuals.  
Additional monitoring or attention may be required for  
patients using this device who are unable to hear or see  
alarms or communicate discomfort.  
DO NOT connect the concentrator in parallel or series with  
other oxygen concentrators or oxygen therapy devices.  
Radio Frequency Interference  
This equipment has been tested and found to comply with  
EMC limits specified by IEC/EN 6060112. These limits are  
designed to provide a reasonable protection against  
electromagnetic interference in a typical medical installation.  
Perfecto2Series  
8
 
   
SECTION 1—GENERAL GUIDELINES  
Other devices may experience interference from even the low  
levels of electromagnetic emissions permitted by the above  
standards. To determine if the emissions from the Perfecto2 is  
causing the interference, turn the Perfecto2 Off. If the  
interference with the other device(s) stops, then the Perfecto2  
is causing the interference. In such rare cases, interference  
may be reduced or corrected by one of the following  
measures:  
• Reposition, relocate, or increase the separation  
between the equipment.  
• Connect the equipment into an outlet on a circuit  
different from that to which the other device(s) is  
connected.  
9
Perfecto2Series  
 
SECTION 2—FEATURES  
SECTION 2—FEATURES  
Oxygen Outlet  
Oxygen Purity Indicator  
Lights / Fault and Power  
Indicator Lights  
Flowmeter  
Circuit Breaker  
Power Switch  
Elapsed Time  
Meter  
ACCESSORIES (NOT SHOWN):  
PreciseRX™ Pediatric Humidifier/Flowmeter  
Accessory - IRCPF16 (Perfecto2 5 Only)  
HomeFill home oxygen compressor - IOH200  
REAR VIEW  
*NOTE: This outlet fitting is to be  
used only for filling oxygen cylinders  
with the HomeFill home oxygen  
compressor. The outlet fitting does not  
affect concentrator performance.  
Refer to the HomeFill ownerʹs  
manual, part number 1100873, for  
connection and operating  
Cabinet Filter  
instructions. When not in use, the  
plug provided with the concentrator  
should be inserted into the outlet  
fitting. For more information about  
the HomeFill, contact your Invacare  
dealer.  
Power  
Cord  
*Outlet  
Fitting  
Perfecto2Series  
10  
 
 
SECTION 3—HANDLING  
SECTION 3—HANDLING  
The concentrator should ALWAYS be kept in the upright  
position to prevent cabinet damage while being transported.  
If the concentrator is to be reshipped by common carrier,  
additional cartons are available from Invacare.  
Unpacking  
NOTE: For this procedure, refer to FIGURE 3.1.  
1. Check for any obvious  
damage to the carton or  
its contents. If damage is  
evident, notify the  
carrier, or your local  
dealer.  
2. Remove all loose packing  
from the carton.  
3. Carefully remove all the  
components from the  
carton.  
FIGURE 3.1 Unpacking  
NOTE: Unless the oxygen concentrator is to be used immediately,  
retain containers and packing materials for storage until use of the  
concentrator is required.  
Inspection  
Inspect/examine exterior of the oxygen concentrator for nicks,  
dents, scratches or other damages. Inspect all components.  
Storage  
1. Store the repackaged oxygen concentrator in a dry area.  
2. DO NOT place objects on top of repackaged concentrator.  
11  
Perfecto2Series  
 
         
SECTION 4—TYPICAL PRODUCT PARAMETERS  
SECTION 4—TYPICAL PRODUCT  
PARAMETERS  
Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60 Hz  
Rated Current Input:  
Sound Level:  
3 A  
43 ± 2 dBA Average for Perfecto2 Models and  
Perfecto2V Model  
40 ± 2 dBA Average for Perfecto2W Models  
Altitude:  
All Perfecto2 Models: Up to 8,000 ft (2438  
meters) above sea level without degredation of  
concentration levels. From 8,000 ft to 13,129  
(4000 meters) below 90% efficiency.  
All Perfecto2 Models while filling a  
HomeFill System: Up to 6,000 ft (1828 meters)  
above sea level without degradation of  
concentration levels. From 6,000 ft (1828 meters)  
to 13,129 ft (4000 meters) below 90% efficiency.  
*Oxygen Output  
87% to 95.6% for flows ranging at 0.5 to 5 L/min  
Concentration Levels:  
Maximum Outlet Pres-  
sure:  
5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)  
Flow Range:  
0.5 to 5 L/min  
For flowrates less than 1 L/min, we recommend  
the use of the Invacare Pediatric Flowmeter  
IRCPF16 Accessory.  
Low Flow Alarm:  
0 L/min to 0.5 L/min  
Rapid Audible Alarm Beeping (No Accessories  
Connected).  
Average Power  
Consumption:  
Perfecto2 and Perfecto2W Models: 280 W  
(Operating @ 3 L/min)  
Perfecto2V Model: 325 W  
Pressure Relief Mecha-  
nism Operational at:  
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)  
Perfecto2Series  
12  
 
   
SECTION 4—TYPICAL PRODUCT PARAMETERS  
Change in maximum  
0.7 L/min  
recommended flow when  
back pressure of 7kPa is  
applied:  
Filters:  
Cabinet, Outlet HEPA and Compressor Inlet  
Safety System:  
Current overload or line surge shutdown. High  
temperature compressor shutdown. High  
Pressure Alarm w/compressor shutdown. Low  
Pressure Alarm w/compressor shutdown.  
Battery Free Power Loss Alarm. SensO2 Oxygen  
System (SensO2 Model) Low Flow Alarm.  
Width:  
Height:  
Depth:  
Weight:  
15 in ± 3/8 in (38.1 cm ± 1 cm)  
23 in ± 3/8 in (58.4 cm ± 1 cm)  
12 in ± 3/8 in (30.5 cm ± 1 cm)  
43 lbs ± 2 lbs (19.5 kg ± 1 kg) Perfecto2 Models  
45 lbs ± 2 lbs (20.5 kg ± 1 kg) Perfecto2W Models  
40 lbs ± 2 lbs (18.1 kg ± 1 kg) Perfecto2V Model  
Shipping Weight:  
48 lbs ± 2 lbs (21.8 kg ± 1 kg) Perfecto2 Models  
50 lbs ± 2 lbs (22.8 kg ± 1 kg) Perfecto2W Models  
45 lbs ± 2 lbs (20.4 kg ± 1 kg) Perfecto2V Model  
Operating Ambient  
Temperature:  
50°F - 95°F (10°C - 35°C) at 20-60% relative  
humidity Perfecto2 Models and Perfecto2V Model  
50°F - 90°F (10°C - 32°C) at 20-60% relative  
humidity Perfecto2W Models  
Exhaust:  
Less than Ambient + 35°F (+19°C)  
Less than Ambient + 6°F (+3°C)  
Oxygen Output  
Temperature:  
Cabinet:  
Impact Resistant flame-retardant plastic cabinet  
that conforms to UL94V-0.  
Regulatory Listing:  
Double Insulated Product ETL certified  
complying with UL1431 and UL1097 ETL certified  
to CSA C22.2 No. 68.  
Electrical:  
No extension cords.  
13  
Perfecto2Series  
 
SECTION 4—TYPICAL PRODUCT PARAMETERS  
Placement:  
Tubing:  
No closer than three inches from any wall,  
furniture, draperies, or similar surfaces.  
7 ft cannula with a maximum 50 ft of Crush-Proof  
Tubing (DO NOT pinch).  
Relative Humidity:  
Time of Operation:  
20 to 60%  
Up to 24 hours per day.  
Recommended Storage  
and Shipping Tempera-  
ture:  
-20°F to 150°F (-29°C to 65°C), at 15-95%  
relative humidity.  
*NOTE: Stated concentration levels achieved after initial warmup  
period (approximately 30 minutes).  
Perfecto2Series  
14  
 
SECTION 5—OPERATING INSTRUCTIONS  
SECTION 5—OPERATING  
INSTRUCTIONS  
Introduction  
Your oxygen concentrator is intended for individual use  
indoors. It is an electronically operated device that separates  
oxygen from room air. It provides high concentration of oxygen  
directly to you through a nasal cannula. Clinical studies have  
documented that oxygen concentrators are therapeutically  
equivalent to other types of oxygen delivery systems.  
Your provider will show you how to use your oxygen  
concentrator. He/She should be contacted with any questions  
or problems regarding your oxygen concentrator. This  
owners manual will tell you about your concentrator and will  
serve as a reference as you use your concentrator.  
Select a Location  
WARNING  
NEVER block the air openings of the product or place it on a  
soft surface, such as a bed or couch, where the air opening may  
be blocked. Keep the openings free from lint, hair and the like.  
Keep unit at least 3 inches away from walls, draperies,  
furniture, and the like.  
You may select a room where using your oxygen concentrator  
would be most convenient. Your concentrator can be easily  
rolled from room to room on its casters.  
Your oxygen concentrator will perform best when operated  
under the following conditions. Refer to Typical Product  
Parameters on page 12. Usage in environments other than  
those described may result in the need for increased  
equipment maintenance. The air intake of the unit should be  
located in a well ventilated area to avoid smoke, airborne  
pollutants and/or fumes. DO NOT place in a closet.  
15  
Perfecto2Series  
 
     
SECTION 5—OPERATING INSTRUCTIONS  
Set Up  
Plug in Power Cord  
Plug in power cord to an electrical outlet.  
Connect Humidifier (If So Prescribed)  
WARNING  
DO NOT overfill humidifier.  
DO NOT reverse the oxygen input and output connections.  
Water from the humidifier bottle will travel through the  
cannula back to the patient.  
NOTE: For this procedure, refer to FIGURE 5.1, FIGURE 5.2 on  
1. Remove cap from bottle.  
2. Fill humidifier with distilled water to the level indicated  
by the manufacturer. Replace the humidifier cap and  
securely tighten.  
Humidifier  
Bottle w/o Cap  
Humidifier  
Bottle with Cap  
FIGURE 5.1 Filling the Humidifier  
3. Insert a flathead screwdriver in the plate groove on the top  
edge of the filter access door and gently pry the filter  
access door off (FIGURE 5.2).  
4. Pull up and remove the humidifier bottle adapter  
Perfecto2Series  
16  
 
   
SECTION 5—OPERATING INSTRUCTIONS  
Humidifier  
Bottle Adapter  
Insert Flat Head  
Screwdriver  
Here  
Cabinet Filter  
Filter Access Door  
FIGURE 5.2 Humidifer Bottle Adapter  
5. Attach the humidifier bottle adapter to the humidifier  
bottle by turning the wing nut on the humidifier bottle  
counterclockwise until it is securely fastened. See Detail  
A” in FIGURE 5.3.  
6. Place the humidifier bottle/adapter assembly in the  
humidifier compartment on the concentrator. See Detail  
“B” in FIGURE 5.3.  
7. Attach oxygen tubing from the humidifier bottle/adapter  
assembly to oxygen outlet connector on the concentrator.  
See Detail “B” in FIGURE 5.3.  
8. Attach the cannula/patient supply tubing to the  
humidifier bottle outlet. See Detail “B” in FIGURE 5.3.  
9. After assembly, ensure that oxygen is flowing through the  
cannula.  
17  
Perfecto2Series  
 
 
SECTION 5—OPERATING INSTRUCTIONS  
Oxygen Outlet  
Connector  
DETAIL “A”  
DETAIL “B”  
Oxygen  
Tubing  
Humidifier  
Bottle  
Humidifier  
Bottle  
Adapter  
Outlet  
Humidifier  
Bottle  
Humidifier  
Bottle  
Humidifier  
Compartment  
FIGURE 5.3 Humidifier Compartment  
Power Switch  
NOTE:Forthisprocedure,referto  
On/Off  
(I/O)  
Power  
Switch  
1. Press power switch to on  
position. All the panel  
lights and the audible  
alarm will come on for  
one second, indicating  
thattheunitisfunctioning  
properly.  
FIGURE 5.4 Power Switch  
Perfecto2Series  
18  
 
     
SECTION 5—OPERATING INSTRUCTIONS  
Flowrate  
NOTE: For this procedure, refer to FIGURE 5.5 on page 20.  
1. Turn the flowrate knob to the setting prescribed by your  
physician or therapist.  
CAUTION  
Never set the flow greater than 5 L/min.  
WARNING  
DO NOT change the L/min setting on the flowmeter unless a  
change has been prescribed by your physician or therapist.  
NOTE: To properly read the flowmeter, locate the prescribed flowrate  
line on the flowmeter. Next, turn the flow knob until the ball rises to  
the line. Now, center the ball on the L/min line prescribed.  
2. If the flowrate on the flowmeter ever falls below 0.5 L/min  
for more than about one minute, the LOW FLOW alarm  
will be triggered. This is a rapid beeping of the audible  
alarm. Check your tubing or accessories for blocked or  
kinked tubing or a defective humidifier bottle. After rated  
flow is restored, between 0.5 L/min and 0.75 L/min, the  
LOW FLOW audible alarm will turn off.  
NOTE: The use of some accessories such as the pediatric flowstand  
and the HomeFill compressor will deactivate the Low Flow Alarm.  
19  
Perfecto2Series  
 
 
SECTION 5—OPERATING INSTRUCTIONS  
Flow Knob  
Flowmeter  
0.5 L/min  
Ball  
Front Panel  
FIGURE 5.5 Flowrate  
Front Panel Indicators - Units with  
SensO2  
The SensO2 feature monitors the purity of oxygen generated  
by the oxygen concentrator. If purity falls below factory preset  
standards, indicator lights on the control panel will  
illuminate.  
OXYGEN PURITY  
Auto Shut Down  
RED  
Use Backup  
Call Supplier  
Below  
Normal  
YELLOW  
Normal  
GREEN  
O2  
Perfecto2Series  
20  
 
   
SECTION 5—OPERATING INSTRUCTIONS  
Initial Startup of the Concentrator  
NOTE: Concentrator may be used during the initial start warmup  
time (approximately 30 min.) while waiting for the O2 purity to  
reach maximum.  
LABEL  
O2 PURITY  
INDICATOR LIGHTS (LED)  
SYMBOL  
SYSTEM OKAY  
O2 over 85% (± 2%)  
GREEN Indicator Light  
O2  
O2 Between 73% (± 3%) YELLOW Indicator light  
A. YELLOW Solid  
B. YELLOW Flashing Sensor  
Failure  
to 85% (±2%)  
Call a qualified technician. Have a  
back-up supply of oxygen ready.  
SYSTEM FAILURE  
RED Indicator Light  
O2 Below 73% (±3%)  
Continuous Audible Alarm  
Sieve-GARD Compressor Shut-  
down. Call a qualified technician.  
Switch to a back-up supply of oxy-  
gen immediately.  
Front Panel Indicators - Units without  
SensO2  
NOTE: For this procedure, refer to FIGURE 5.7.  
Auto Shut Down  
RED  
Use Backup  
Call Supplier  
Normal  
I/O  
GREEN  
21  
Perfecto2Series  
 
   
SECTION 5—OPERATING INSTRUCTIONS  
Initial Startup of the Concentrator  
NOTE: Concentrator may be used during the initial start warmup  
time (approximately 30 min). O2 purity will build to a maximum  
during this period.  
LABEL  
SYMBOL  
STATUS  
INDICATOR LIGHTS  
(LED)  
SYSTEM OKAY  
GREEN Indicator Light  
I/O  
SYSTEM FAILURE  
RED Indicator light  
Continuous Audible Alarm  
Sieve-GARD™Compressor  
Shutdown  
Call a qualified technician. Switch  
to a back-up supply of oxygen  
immediately.  
Perfecto2Series  
22  
 
SECTION 6—MAINTENANCE  
SECTION 6—MAINTENANCE  
WARNING  
The Invacare concentrators are specifically designed to  
minimize routine preventive maintenance. Only qualified  
personnel should perform preventive maintenance on the  
concentrator.  
Power should be disconnected before beginning preventive  
maintenance on the Invacare concentrator.  
NOTE: At a minimum, preventive maintenance MUST be  
performed according to the maintenance record guidelines. In places  
with high dust or soot levels, maintenance may need to be performed  
more often. Refer to Preventive Maintenance Record on page 25.  
Routine Maintenance  
NOTE: The following routine maintenance should be performed  
more frequently. See recommended intervals in each section and  
preventive maintenance record.  
Cleaning the Cabinet Filter  
CAUTION  
DO NOT operate the concentrator without the filter installed.  
NOTE: For this procedure, refer to FIGURE 6.1 on page 24.  
NOTE: There is one cabinet filter located on the back of the cabinet.  
1. Remove the filter and clean at least once a week depending  
on environmental conditions.  
NOTE: Environmental conditions that may require more frequent  
cleaning of the filters include but are not limited to: high dust, air  
pollutants, etc.  
23  
Perfecto2Series  
 
     
SECTION 6—MAINTENANCE  
2. Clean the cabinet filter with a vacuum cleaner or wash in  
warm soapy water and rinse thoroughly.  
3. Dry the filter thoroughly before reinstallation.  
Cleaning the Cabinet  
Clean the cabinet with a mild household cleaner and  
nonabrasive cloth or sponge.  
Cleaning the Humidifier  
NOTE: To clean the oxygen humidifier, follow the instructions  
provided by the manufacturer. If none are provided, follow these  
STEPS:  
1. Clean the humidifier every day.  
2. Wash it in soapy water and rinse it with a solution of ten  
parts water and one part vinegar.  
3. Rinse thoroughly with hot water and refill with distilled  
water to the level shown on the humidifier.  
Perfecto2Series  
24  
 
 
SECTION 6—MAINTENANCE  
MODEL NO. IRC  
SERIAL NO.  
FIGURE 6.2 Preventive Maintenance Record  
25  
Perfecto2Series  
 
SECTION 7—TROUBLESHOOTING GUIDE  
SECTION 7—TROUBLESHOOTING  
GUIDE  
SYMPTOM:  
PROBABLE CAUSE:  
SOLUTION:  
Alarm:  
Main Power Loss:  
1. Insert plug into outlet.  
Short beeps, long 1. Power cord not  
pause  
plugged in.  
2. No power at outlet.  
2. Inspect house circuit  
breakers or fuses. If problem  
recurs, use a different outlet.  
Concentrator not  
operating,  
power switch On.  
3. Tripped circuit  
breaker.  
3. Call Service Provider.  
Beep......Beep......  
Alarm:  
Continuous  
System Failure:  
1. Unit overheating due  
to blocked air intake.  
1a. Remove and clean cabinet  
filters.  
1b. Move oxygen  
Concentrator not  
operating, Power  
Switch On.  
concentrator at least three  
inches away from walls,  
draperies or furniture.  
2. Insufficient power at  
outlet.  
2. DO NOT use extension  
cords. Move to another  
electrical outlet or circuit.  
Beeeeeeeep....  
3. Internal repairs  
required.  
3. Call Service Provider.  
Alarm:  
Rapid  
Low Flow Alarm:  
1. Kinked or blocked  
tubing, cannula or  
humidifier.  
1. Inspect for kinks or  
blockages. Correct, clean or  
replace item. Once  
corrected, turn power Off for  
60 seconds and then turn  
power back ON.  
Beep....Beep...  
Beep...Beep  
2. Flowmeter set at 0.5  
L/min or less.  
2. Reset flowmeter to  
prescribed flowrate. If  
prescribed flowrate is below  
0.5 L/min, use the Pediatric  
Flowmeter IRCPF16 (Only  
with Perfecto2 5).  
Perfecto2Series  
26  
 
 
SECTION 7—TROUBLESHOOTING GUIDE  
SYMPTOM:  
PROBABLE CAUSE:  
SOLUTION:  
YELLOW or RED 1. Low oxygen purity.*  
1. Clean or Replace filters.  
light Illuminated.  
2. Kinked or blocked  
tubing, cannula or  
humidifier.*  
2. Inspect for kinks or  
blockages. Correct, clean or  
replace item. Once  
corrected, turn power Off for  
60 seconds and then turn  
power back On.  
Alarm:  
Continuous  
On RED light  
only.  
3. Flowmeter set at 0.5  
* Only applies to  
IRC5PO2.  
3. Check flowmeter is set to  
1.0 L/min or more.  
L/min or less.*  
4. Unit overheating due  
to blocked air intake.  
4a. Remove and clean cabinet  
filters.  
4b. Move concentrator at  
least three inches from walls,  
draperies, and furniture.  
5. Insufficient power at  
outlet.  
5. DO NOT use extension  
cords. Move to another  
electrical outlet or circuit.  
6. Internal repairs  
required.  
6. Call Service Provider.  
GREEN light with 1. Internal repairs  
1. Call Service Provider.  
YELLOW Light  
Flashing.  
required.  
27  
Perfecto2Series  
 
NOTES  
NOTES  
Perfecto2Series  
28  
 
NOTES  
NOTES  
29  
Perfecto2Series  
 
NOTES  
NOTES  
Perfecto2Series  
30  
 
LIMITED WARRANTY  
LIMITED WARRANTY  
NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY  
WITH FEDERAL LAW APPLICABLE TO PRODUCTS  
MANUFACTURED AFTER JULY 4, 1975.  
This warranty is extended only to the original purchaser who purchases  
this product when new and unused from Invacare Corporation or a  
dealer. This warranty is not extended to any other person or entity and it  
is not transferable or assignable to any subsequent purchaser or owner.  
Coverage under this warranty will end upon any such subsequent sale or  
other transfer of title to any other person. This warranty gives you specific  
legal rights and you may also have other legal rights which may vary from  
state to state.  
Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when  
purchased new and unused to be free from defects in materials and  
workmanship for a period of three years from date of purchase from  
Invacare or a dealer, with a copy of the seller’s invoice required for  
coverage under this warranty.  
DO NOT OPEN OR ATTEMPT TO SERVICE THE 4-WAY VALVE  
BODY. This will void any and all warranty applicable to the valve body. If  
within such warranty periods any such product shall be proven to Invacare  
Corporation's satisfaction to be defective, such product shall be repaired  
or replaced, at Invacare Corporation's option. This warranty only applies  
to the labor for repairs performed by the Invacare Service Department or  
Invacare Authorized Service Centers. It does not apply to the labor  
performed by the purchaser or user. This warranty does not include  
normal wear and tear or shipping charges. Invacare Corporation's sole  
obligation and your exclusive remedy under this warranty shall be limited  
to such repair or replacement. Routine maintenance items, such as filters,  
are excluded from this warranty.  
For warranty service, please contact Invacare Corporation's Service  
Department at the toll free number on the back cover. Upon receiving  
notice of an alleged defect in a product, Invacare Corporation will issue a  
serialized return authorization. It shall then be the responsibility of the  
purchaser to return the entire unit or remove, at purchaser's cost, the  
defective component part(s) identified, pack the component part(s) in a  
manner to avoid shipping damage and to ship the component part(s) to  
either Invacare Corporation's plant or service center as specified by  
Invacare Corporation in advance. Defective component part(s) MUST be  
returned for warranty inspection using the serial number as identification  
within thirty days of return authorization date. DO NOT return products  
to our factory without prior consent. C.O.D. shipments will be refused;  
please prepay shipping charges.  
31  
Perfecto2Series  
 
 
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY  
SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE,  
ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE,  
SOOT OR SMOKE-FILLED ENVIRONMENTS, OR OTHER THAN  
NORMAL APPLICATION, USE OR SERVICE, OR TO PRODUCTS  
MODIFIED WITHOUT INVACARE CORPORATION'S EXPRESS  
WRITTEN CONSENT (INCLUDING, BUT NOT LIMITED TO,  
MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS  
OR ATTACHMENTS) OR TO PRODUCTS DAMAGED BY REASON OF  
REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC  
CONSENT OF INVACARE CORPORATION OR TO PRODUCTS  
DAMAGED BY CIRCUMSTANCES BEYOND INVACARE  
CORPORATION'S CONTROL.  
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN  
LIEU OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER  
EXPRESS OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF  
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE,  
AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY  
WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT  
OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS  
CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED  
WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE  
DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN.  
INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR  
INCIDENTAL DAMAGES WHATSOEVER.  
SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION  
OF INCIDENTAL OR CONSEQUENTIAL DAMAGE, OR LIMITATION  
OF HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE  
EXCLUSION AND LIMITATION MAY NOT APPLY TO YOU.  
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH  
STATE/PROVINCIAL LAWS AND REQUIREMENTS.  
Invacare Corporation  
USA  
© 2009 Invacare Corporation. All rights  
reserved. Republication, duplication or  
modification in whole or in part is  
prohibited without prior written  
permission from Invacare. Trademarks are  
identified by ™ and ®. All trademarks are  
owned by or licensed to Invacare  
Corporation or its subsidiaries unless  
otherwise noted.  
One Invacare Way  
Elyria, Ohio USA  
44036-2125  
800-333-6900  
Technical Services  
800-832-4707  
Part No 1143482  
Rev E - 11/09  
 
   

Indesit Washer SIXL 145K User Manual
Invacare Mobility Aid TerminatorTM SS User Manual
Jenn Air Dishwasher JDB1270AWF User Manual
JVC Computer Monitor KW AVX700 User Manual
Kalorik Toaster 33005 User Manual
Karcher Pressure Washer K205 M Plus User Manual
Kenmore Gas Heater 15333096 User Manual
Kenwood Car Amplifier TL 922A User Manual
Kenwood Stereo Receiver KR V888D User Manual
Keys Fitness Fitness Equipment KF 45 User Manual